UT4000 User Manual
Chapter 6 Appendixes
Appendix A : Technical Indicators
Unless otherwise specified, all technical specifications apply to probes with a attenuation
setting of 10X and the UT4000 Series Digital Oscilloscopes. To verify that your
oscilloscope meets specifications, it must first meet the following conditions :
The oscilloscope must have been operating continuously for thirty minutes within
the specified operating temperature.
If the operating temperature changes by more than 5°C, you must perform the “Self
Cal” operation, accessible through the System Functions menu. All specifications
are guaranteed unless noted “Typical”.
Technical Indicators
Acquisition
Acquisition modes Real time Equivalent
Acquisition rates 2GS/s 50GS/s
(UT4302C/UT4202C/UT4152C)
Average When all channels have made N acquisitions simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 to 256.
Input
Input coupling DC, AC, GND
Input impedance 1M±1.5% in parallel with 24pF±3pF.
Probe attenuation factor 1X, 10X, 100X, 1000X.
Maximum input voltage 400V (DC + AC Peak, 1 M input impedance)
(10:1 probe attenuation); 5V (Vrms, BNC).
Time delay between channels (Typical) 150ps